Laminar Airflow cabinets are designed for creation of a bacterial dust free air space. Laminar Airflow Cabinets are used for work with low-risk substances and materials, when protection of working material from environment is required or work with item requires a sterile working zone. Cabinets are used for equipment of local workplaces in medical, pharmaceutical and other institutes with high requirements for air cleanliness in the working zone.
|
 | Horizontal Laminar Air Flow | | |
| We deal in manufacturing of horizontal laminar air flow that finds it applications in various hospitals, electronic industry and others. We use high grade component and quality raw material that is tested on various quality parameters and ensure smooth functionality.
Specifications: - Direction of flow: Horizontal
- Cleanliness: CLASS 100
- Velocity: 90 FPM + / -20%
- Hepa Filter: Size and QTY. will depend on the Size selected
- Particle retention: Above 0. 3 micron
- Blower Assembly: 1 / 3 HP, 1440 RPM
- Noise level: 65 decibel + / -5%
- Rate of air flow: 0. 40 to 0. 75 M / Sec. (90 FPM)
- Vibration level: < 2. 5 micron
|
 | Vertical Laminar Air Flow | | |
Our vertical laminar air flow has vertical direction flow and particle retention above 0.5 micron. These are appreciated for their low noise and versatile usage. These have been designed with latest technology and can also be designed as per the specifications, provided by our clients.
Application : The vertical laminar air flow cabinets is designed to provide a high degree of protection for process products In laboratories and production facilities. Many critical applications in the medical, pharmaceutical, nuclear-power, and micro electronic fields demand an ultra-clean work environment which is free from biological and particulate contamination.
Operation : Ambient Air is draw in through pre-filter and is introduced in the work zone through the HEPA filter. The average velocity in the work zone is maintained between 0.4 and 0.5 m/s. Air cleanliness in the work zone is better than class 3.5 (100) in accordance with US FED STD 209 E
Construction : The cabinets are fabricated out of thick board duly sunmica clad or mild steel cabinet duly epoxy coated. The work table is made of thick board which is sunmica clad at top or S. S Table top will be provided as optional accessories. All joints are sealed with silastic sealant. Side panels are provided of thick transparent acrylic sheets. The unit is fitted with pre-filter and is made to pass through highly effective HEPA (High Efficiency Particular Air) filter having efficiency rating as high as 99.997% with hot, DOP, thus retaining all air-borne particles of size : 0.3 micron and larger. Using a dynamic machine, the blower and motor assembly is statically and dynamically balanced. Blowers are mounted on vibration isolation mounts to minimize vibration. The working is illuminated by fluorescent lightings fitted to the unit. Height of the working table provides comfortable “ Sit Down” working position for the operation to work on 220/230 volts AC. supply.
| Air Velocity | Between 0.4 to 0.5 mls +-20% | | HEPA Efficiency Pre-filter | 99.997% at 0.3 microns with DOP test Washable with an arrestance up to 90% at 5 microns Less than 65 dB (A) | | Sound Level | Lux 900 - 1300 Lux | | Light Intensity Cleanliness | Standard 209 E Class 100 | | Federal Power Supply | 220 V Single phase 50 Hz | Optional : • S. S Table Top. • U. V Lights attachment • Front Acrylic Door • Gas/ Vacuum Cock • Castor wheels at the Bottom | VERTICAL LAMINAR FLOW | HORIZONTAL LAMINAR FLOW | | MODEL | SIZE IN FT | MODEL | SIZE IN FT | | VL-2A | 2X2 | HL-2A | 2X2 | | VL-3B | 3X2 | HL-3B | 3X2 | | VL-4C | 4X2 | HL-4C | 4X2 | | |
 | Ceiling Laminar Airflow | | |
| Ceiling suspended laminar air flow (CLAF) are used for OT applications, pharma filling, operation theatre & packing line etc. CLAF provides ultra clean sterile HEPA filtered CLASS 100 air in a isolated area where the highly sensitive process activities are carried-out. Cleanair CLAF Laminar Air Flow available in complete stainless steel construction or GI construction with PU coated finish. Equipped with MINIPLEAT HEPA with dynamically balanced ultra-low noise blowers for optimum airflow. Differential pressure monitoring done through MAGNEHELIC gauges. Various sizes can be manufactured and supplied to suit the specific site conditions. |
 |
Industry We Cater to |
|
|
Biotechnology Tissue culture Bio Chemistry Botany Zoology Chemistry Biology Hospitals Diagnostics Labs Pharma Industry Healthcare Industry Vaccine manufacturers Food Processing industry Electronics service centre Semiconductors Space Applications Life Science Chemical Industry Bio medical Micro electronics Micro biology Clean room Blood bank College – zoology , botany , chemistry Medical implant Medical devices Educational institutions
|
|
 Product Showroom View all our products / services as a slide show. play now

|